Zhongke Meiling Cryogenics Co., Ltd
Meling Biomedical YC-650L 2°C to 8°C Upright Medical Refrigerator, 22.95 cu. ft. (650 L)
Meling Biomedical YC-650L 2°C to 8°C Upright Medical Refrigerator, 22.95 cu. ft. (650 L)
Couldn't load pickup availability
Selecting the YC-650L
Solution Instruments supplies the Meling YC-650L in Canada and the United States. Compare this 650 L upright laboratory refrigerator for 2°C to 8°C storage using its five shelves. Include the required temperature records, any remote monitoring and any site mapping or qualification documentation in the quotation scope.
Request a YC-650L quoteCompare 2–8°C refrigerators
meling-biomedical-yc-650l-specification-sheet.pdf · Authorized distributor information
The YC-650L provides 650 L (22.95 cu. ft.) of refrigerated storage with a 2°C to 8°C temperature range. Air cooling, automatic defrost and a digital display support everyday refrigerated storage.
The cabinet measures 715 × 941 × 1,985 mm (28.15 × 37.05 × 78.15 in), with five plastic-coated steel wire shelves in a stainless-steel interior.
Certifications and Regulatory Status
The 110 V / 60 Hz YC-650L has the following certifications and regulatory status:
- UL/cUL listed
- FDA Facility Registration and Product Listing
Temperature Control and Refrigeration
The YC-650L uses microprocessor control, air cooling, automatic defrost and one compressor with R290 (propane) refrigerant. Display temperature accuracy is 0.1°C. The operating ambient temperature range is 16°C to 32°C.
The following performance values apply at a 5°C setpoint and 25°C ambient temperature:
- Temperature uniformity: 2.2°C
- Temperature fluctuation: 1.8°C
- Pull-down time to 5°C: 28 minutes
- Temperature recovery after a 60-second door opening: 7.5 minutes
- Power consumption: 5 kWh/24 h
- Noise level: 49.58 dB(A)
Temperature Monitoring and Interfaces
The built-in data logger records temperature every 10 minutes and stores up to 100,000 data records. Records can be exported via USB.
Battery backup duration: 36–48 hours for the temperature display and alarms during a power failure.
An optional Monitor system provides remote high/low temperature alarms and viewing of temperature history over 24 hours.
Alarm Functions
The YC-650L includes alarms for:
- High and low storage temperatures
- High ambient temperature
- High condenser temperature
- Power failure
- Low battery
- Sensor error
- Data communication failure
- Door ajar
- Data logger failure
Cabinet Features
- Pre-coated metal (PCM) exterior
- Stainless-steel interior
- Double-pane glass door
- Ergonomic handle
- Five plastic-coated steel wire shelves with adjustable heights
- Polyurethane (PU) insulation
- One access test port, 23 mm in diameter
- Four casters, including two with brakes
Electrical Specifications
| Power supply | Rated current at 32°C ambient |
|---|---|
| 110 V / 60 Hz | 5.77 A |
Main Specifications
| Specification | YC-650L |
|---|---|
| Capacity | 650 L / 22.95 cu. ft. |
| Temperature range | 2°C to 8°C |
| Operating ambient temperature | 16°C to 32°C |
| Display | Digital |
| Display temperature accuracy | 0.1°C |
| Cooling method | Air cooling |
| Defrost method | Automatic |
| Number of compressors | 1 |
| Refrigerant | R290 (propane) |
| Exterior material | Pre-coated metal (PCM) |
| Interior material | Stainless steel |
| Shelves | 5 |
| Shelf material | Plastic-coated steel wire |
| Insulation material | Polyurethane (PU) |
| Insulation thickness | Door: 55 mm; left/right sides: 55 mm; back: 55 mm; top: 55 mm; bottom: 62 mm |
| External dimensions, width × depth × height | 715 × 941 × 1,985 mm / 28.15 × 37.05 × 78.15 in |
| Internal dimensions, width × depth × height | 605 × 725 × 1,515 mm / 23.82 × 28.54 × 59.65 in |
| Package dimensions, width × depth × height | 773 × 947 × 2,154 mm / 30.43 × 37.28 × 84.80 in |
| Net weight | 157 kg / 346.13 lb |
| Gross weight | 185 kg / 407.86 lb |
| Access test port | 1 port, 23 mm diameter |
| Casters / brakes | 4 / 2 |
| Data logging / export | Every 10 minutes / 100,000 records / USB export |
| Backup battery — display and alarms | 36–48 h |
| Monitor system | Optional; remote high/low temperature alarms and 24-hour temperature history |
| Certifications and regulatory status | UL/cUL listed; FDA Facility Registration and Product Listing |
Model data for equipment qualification
Meling reports temperature uniformity of 2.2°C and fluctuation of 1.8°C at a 5°C setpoint and 25°C ambient. The built-in logger records every 10 minutes and supports USB export.
If your site requires temperature mapping or installation and operational qualification, include the acceptance criteria in your request so we can confirm available model documentation. Request a YC-650L quote.
Technical documents
Share