The Meling Biomedical DW-FL270 provides 270 L (9.53 cu. ft.) of frozen storage within an upright cabinet. Its −20°C to −40°C temperature range, microprocessor controller, temperature logging and multiple alarm functions support temperature-controlled storage in laboratory and biomedical environments.
Seven ABS drawers help organize stored materials. The cabinet measures 700 × 640 × 1,792 mm (W × D × H) and features a solid door, safety lock, circuit breaker, four casters and two levelling feet.
Certifications
- UL/cUL listed for the United States and Canada
- CE marked
Temperature Control and Refrigeration
The DW-FL270 uses a microprocessor controller and digital display, with temperature readings displayed in 0.1°C increments. Its refrigeration system features direct cooling, manual defrosting, one compressor and R290 hydrocarbon refrigerant.
Manufacturer internal testing reports the following typical performance:
- Temperature uniformity: 3.4°C
- Pull-down time to −40°C: 2.4 hours
- Temperature recovery after a one-minute door opening: 0.6 hours, equivalent to 36 minutes
These figures are based on internal testing at a −40°C set point and a 22°C ambient temperature.
Temperature Logging and Monitoring
The built-in data logger records temperature every 10 minutes and stores up to 100,000 temperature records. A USB port supports data export.
An RS485 interface and remote alarm contact provide additional connection options. The freezer also includes a backup battery and one 25 mm-diameter access port.
Alarm Functions
The DW-FL270 provides alarms for:
- High and low storage temperatures
- High ambient temperature
- Power failure
- Low battery
- Data logger failure
- Sensor error
- Data communication failure
- Door ajar
Cabinet Features
- Seven ABS storage drawers
- Solid door
- Safety lock
- PCM exterior
- Sprayed aluminum plate interior, with stainless steel available as an option
- Circuit breaker for electrical protection
- One 25 mm-diameter access port
- Four casters and two levelling feet
Electrical Specifications
The manufacturer’s brochure lists the following paired power supply and rated current values:
| Power supply (V/Hz) | Rated current (A) |
|---|---|
| 110 V / 60 Hz | 4.22 A |
| 220–240 V / 50 Hz | 2.8 A |
| 220–240 V / 60 Hz | 2.22 A |
Main Specifications
| Specification | DW-FL270 |
|---|---|
| Capacity | 270 L / 9.53 cu. ft. |
| Temperature range | −20°C to −40°C |
| Cooling performance | −40°C |
| Ambient temperature range | 16°C to 32°C |
| Climate class | N |
| Controller | Microprocessor |
| Display | Digital display, with 0.1°C temperature increments |
| Temperature uniformity | 3.4°C |
| Pull-down time to −40°C | 2.4 hours |
| Temperature recovery after a one-minute door opening | 0.6 hours / 36 minutes |
| Cooling method | Direct cooling |
| Defrost mode | Manual |
| Number of compressors | 1 |
| Refrigerant | R290 |
| Insulation thickness | D: 90 mm; L/R: 100 mm; B: 100 mm |
| Exterior material | PCM |
| Interior material | Sprayed aluminum plate; stainless steel optional |
| Drawers | 7 ABS drawers |
| Internal dimensions, W × D × H | 500 × 460 × 1,235 mm / 19.69 × 18.11 × 48.62 in |
| External dimensions, W × D × H | 700 × 640 × 1,792 mm / 27.56 × 25.2 × 70.55 in |
| Package dimensions, W × D × H | 835 × 775 × 1,910 mm / 32.9 × 30.5 × 75.2 in |
| Net weight | 98 kg |
| Gross weight | 115 kg |
| Door | Solid door |
| Safety lock | Included |
| Circuit breaker | Included |
| Access port | 1, diameter 25 mm |
| Mobility | 4 casters and 2 balancing feet |
| Temperature recording interval | Every 10 minutes |
| Data storage capacity | Up to 100,000 temperature records |
| Data export | USB |
| Interfaces | RS485 and remote alarm contact |
| Backup battery | Included |
| Certifications and markings | UL/cUL listed; CE marked |